NanoDrop™ Reagents
NanoDrop™ Reagents
Actual product may vary
NanoDrop™ Reagents
NanoDrop™ Reagents
NanoDrop™ Reagents
NanoDrop™ Reagents
Thermo Scientific™

NanoDrop™ Reagents

Certified standards for verifying instrument performance to factory specifications are available separately.
Have Questions?
Change viewbuttonViewtableView
Catalog NumberDescription
CHEM-PR1-KITNanoDrop PR-1 Reconditioning Kit
CHEM-CF-1NanoDrop CF-1 Calibration Check Fluid
CHEM-CF-8CF-8 Calibration Check Kit
CHEM-PV-1NanoDrop PV-1 Performance Verification Solution
CHEM-PV-8NanoDrop PV-8 Performance Verification Kit
Catalog number CHEM-PR1-KIT
Price (USD)
139.00
Each
Add to cart
Description:
NanoDrop PR-1 Reconditioning Kit
Price (USD)
139.00
Each
Add to cart
Certified standards for verifying instrument performance to factory specifications are available separately.

CHEM-PV-1 - PV-1 Performance Verification Solution

For use with NanoDrop One/OneC and NanoDrop Lite Plus
Solution with certified absorbance values.
Use to create a record of correct photometric performance.
Intended for single use only.
 

CHEM-PV-8 - PV-8 Performance Verification Kit

For use with NanoDrop Eight
Solution with certified absorbance values.
Use to create a record of correct photometric performance.
Intended for single use only.
 

CHEM-CF-1 - CF-1 Calibration Fluid

For use with NanoDrop 1000, NanoDrop 2000/2000c, NanoDrop Lite
Solution with certified absorbance values.
Use to create a record of correct photometric performance.
Intended for single use only.
 

CHEM-CF-8 - CF-8 Calibration Check Kit

For use with NanoDrop 8000
Solution with certified absorbance values.
Use to create a record of correct photometric performance.
Intended for single use only.
 

CHEM-PR1-KIT - PR-1 Reconditioning Kit

For use with all NanoDrop models
This kit is useful for reconditioning the sample measurement pedestals to a hydrophobic state if they become “unconditioned” (see user manual for more information). The PR-1 kit consists of a small screw cap vial of specially formulated compound and a supply of 25 convenient applicators.

Specifications
DescriptionNanoDrop PR-1 Reconditioning Kit
TypeNanoDrop PR-1 Reconditioning Kit
For Use With (Equipment)All Thermo Scientific NanoDrop Spectrophotometers
Unit SizeEach

Frequently asked questions (FAQs)

How do I keep my sample from spreading on the pedestal of my NanoDrop Lite Spectrophotometer?

There are many reagents and organic solvents that could compromise the pedestal surface properties, causing samples to flatten out rather than bead up. Use the NanoDrop PR-1 reconditioning compound as a rapid means of reconditioning the pedestals when the pedestal surface properties have been compromised and samples spread out on the pedestal. NanoDrop PR-1 Reconditioning Kits are available through Thermo Fisher website (Cat. No. CHEM-PR1-KIT)

My NanoDrop 8000 Spectrophotometer failed calibration check. What should I do?

Please review the following steps to take when/if your NanoDrop 8000 Spectrophotometer fails a calibration check:
- Clean the pedestal with the NanoDrop PR-1 Reconditioning Kit
- Repeat calibration check using a FRESH vial of NanoDrop CF-1 Calibration Fluid and FRESH blanking solution
- If instrument continues to be out of specification, Contact Technical Support

My NanoDrop 2000/2000c Spectrophotometer failed the calibration check. What should I do?

Please review the following steps to take when/if your NanoDrop 2000/2000c fails a calibration check:
- Clean the pedestal with the NanoDrop PR-1 Reconditioning Kit
- Repeat calibration check using a FRESH vial of NanoDrop CF-1 Calibration Fluid and FRESH blanking solution
More detailed instructions can be found on the Calibration Check Procedure document: https://assets.thermofisher.com/TFS-Assets/CAD/Product-Bulletins/NanoDrop-2000-2000c-1000-Calibration-Check-procedure.pdf
If your NanoDrop 2000/2000c continues to be out of specification, Contact Technical Support.