TSQ Quantis™ Triple Quadrupole Mass Spectrometer
TSQ Quantis™ Triple Quadrupole Mass Spectrometer
TSQ Quantis™ Triple Quadrupole Mass Spectrometer
TSQ Quantis™ Triple Quadrupole Mass Spectrometer
TSQ Quantis™ Triple Quadrupole Mass Spectrometer
TSQ Quantis™ Triple Quadrupole Mass Spectrometer
Thermo Scientific™

TSQ Quantis™ Triple Quadrupole Mass Spectrometer

Perform robust targeted quantitation workflows with best-in-class sensitivity for all molecule types and user expertise levels.

Have Questions?
Catalog NumberDescription
TSQ02-10001TSQ Quantis triple-stage quadrupole mass spectrometer
Catalog number TSQ02-10001
Price (USD)
-
Request A Quote
Description:
TSQ Quantis triple-stage quadrupole mass spectrometer

Easily answer critical analytical challenges during targeted quantitation workflows in your laboratory. With its segmented quadrupoles, faster rod drivers, and improved electron multipliers, the Thermo Scientific™ TSQ Quantis™ Triple Quadrupole Mass Spectrometer delivers unprecedented robustness with best-in-class sensitivity for everyday use and for every user.

The TSQ Quantis MS seamlessly integrates with industry-leading quantitation software, HPLCs and UHPLCs, and sample preparation solutions; it also offers enhanced efficiency during targeted quantitation workflows for a wide array of molecular species. Other benefits include:

  • Superb sensitivity that generates data for all molecular species, even in complex matrices.
  • Outstanding robustness that enables increased confidence in data while prolonging instrument uptime.
  • Reliability and reproducibility improve data quality for every run and every sample.
  • Close integration with application-specific software ensures increased productivity across all application markets.
  • Simplicity and ease-of-use allow users of all expertise levels to acquire high quality data with improved confidence in results.
Specifications
TypeTriple Quadrupole LC-MS
Mass Range5 to 3000 m/z
DescriptionTSQ Quantis triple-stage quadrupole mass spectrometer
For Use With (Application)Pharmaceutical, Environmental, Food Safety, Clinical Research, and Forensic Toxicology
Unit SizeEach

Frequently asked questions (FAQs)

How do I clean/store my Triple Quadrupole mass spectrometry instruments after analysis?

For short periods (less than 24 hr) the system can just be put into standby. However, it is good practice to flush the system with 50:50 solvent/water (where the solvent is either methanol or acetonitrile) to remove buffers or additives and stop microbial growth.

For longer periods, flush the LC system with 50:50 solvent/water. Then flush with 100% solvent. Put the system into standby.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

How do I perform mass calibration on Triple Quadrupole mass spectrometry instruments and how often should I do it?

The mass calibration should ideally be performed every 3 months and definitely performed at least every 6 months. Refer to the Instrument Manual (https://tools.thermofisher.com/content/sfs/manuals/TSQ-Series-Hardware.pdf) for detailed instructions on how to perform mass calibration.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

How should I clean the Ion Transfer Tube on Triple Quadrupole mass spectrometry instruments?

The Ion Transfer Tube should be removed and sonicated in a 50:50 solution of methanol/water that contains 20% formic acid for 30 min. Rinse thoroughly with water. Then sonicate in methanol for 15 min. Dry with nitrogen gas.

The same procedure can be applied to the Ion Sweep Cone.

Refer to the Instrument Manual (https://tools.thermofisher.com/content/sfs/manuals/TSQ-Series-Hardware.pdf) for more detailed instructions.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

How should I adjust my ion source position on Triple Quadrupole mass spectrometry instruments?

Perform loop injections of your analyte using isocractic elution through a short column at the flow rate and approximate mobile phase composition. Use a concentration where the signal to noise is approximately 10:1. Record the raw data file and integrate the peaks, comparing peak areas at different source positions (i.e., left and right, up and down, fore and aft). Ensure that the background noise is not increasing more than the peak area.

Note: The needle position on the Thermo Fisher Scientific TSQ Altis MS, TSQ Quantis MS, and TSQ Fortis MS instruments is set and should not be adjusted.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

Do I need to warm/equilibrate Triple Quadrupole mass spectrometry instruments prior to starting a session of analysis?

It is advisable to turn both the LC system and the MS system on for at least 30 min prior to analysis. This time can usually be spent testing one or more solvents or matrix blanks.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.